Dental Burs USER GUIDE:
■ Ensure that the correct bur is selected, (shape, size and grit type) according to the preparation being carried out.
■ Ensure that all handpieces, high speed and contra-angles, are in perfect working order .
■ Ensure that the bur’s shank is fully inserted and adequately gripped by the handpiece.
■ Ensure that you have selected the correct speed to ensure optimum performance and strictly obser ve the recommended speed constraints as shown below. Only apply the diamond grit to the surface, once the recommended operational speed has been reached.
	■  Ensure  that   preparation work    is   carried out without exerting excess  pressure. The recommended               contact pressure      range  is
	0.3–2N.  Excess  pressure will adversely affect the outcome of your preparation.
■ Ensure that you use an adequate supply of water in order to prevent damaging the tooth structure, pulp and any adjacent fillings. To achieve this, Crosstech recommends high - speed handpieces with three coolant spray jets, which spray the coolant along the entire length of the instrument’s cutting surface. The minimum coolant supply is 50ml/min. Note that burs with lengths in excess of 21mm and a head exceeding 2mm in diameter require additional cooling.
■ Avoid using blunt, damaged, bent or non - concentric diamond instruments. They will adversely affect your preparation work and can damage the handpiece turbine.
| Dental Burs R e c o m m e n d e d S p e e d s | ||
| 
						Head Diamet | 
						er Speed Range | 
						Max. Speed | 
| 008–010 | 75,000–150,000 | 450,000 | 
| 012–014 | 60,000–110,000 | 450,000 | 
| 016–018 | 45,000–88,000 | 450,000 | 
| 021–023 | 40,000–75,000 | 300,000 | 
| 025–027 | 30,000–65,000 | 160,000 | 
| 029–031 | 25,000–56,000 | 140,000 | 
| 033–040 | 22,000–45,000 | 120,000 | 
| 042–050 | 20,000–37,000 | 95,000 | 
| 055–065 | 17,000–32,000 | 80,000 | 
	■ All diamond burs must be disinfected and sterilized prior to use.
	■ After use, diamond instruments must be immediately immersed in a suitable cleaning agent. Use  a cleaning  agent  that contains  a corrosion inhibitor. Do not use strong acidic or alkaline disinfectants  as these will adversely affect  the  instruments.   If  necessar y,  clean by mechanical  means  such  as  with a brass bristled brush ensuring that absolutely no residue is left.
	■  Crosstech diamond   burs  may   be  cleaned ultrasonically  and  sterilized  in an autoclave, chemiclave or in hot air .
	■  After sterilization, keep burs in our dust free bur blocks ready for re-use.
Important: To prevent infectious germs from spreading, please carefully dispose of instruments contaminated with blood or saliva.
| G R I T | I S O N O . | C O D E | µ m | 
| ■ Standard | 524 | 
 | 105–125 | 
| ■ Coarse | 534 | C | 150 | 
| ■ Fine | 514 | F | 55 | 
| ■ Extra Fine | 504 | P | 25 |